Catalog Number RC421C SMN10444986 |
Device Problems
Use of Device Problem (1670); Device Issue (2379)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 07/05/2016 |
Event Type
malfunction
|
Manufacturer Narrative
|
The cause of the biased low qc and patient results is user error.The root cause of the issue was the customer was using an inappropriate calibrator for the dimension exl troponin i method.The customer had been calibrating the dimension exl tni method with an incorrect calibrator (dimension clinical chemistry system heterogenous immunoassay module ctni calibrator) rather than the correct calibrator (dimension exl loci troponin i calibrator).Patients results in the instrument data that had been calibrated with the incorrect calibrator were recalculated.Only one sample showed a result above 0.056 ng/ml that was reported with tni of <0.056 ng/ml.The sample id was (b)(6) from (b)(6) and a 0.048 ng/ml was reported from the rxl calibration but it is showing as a 0.060 ng/ml with a loci calibrator calibration.The dimension exl troponin i method instructions for use indicates that the range encompassing the 99th percentile of normal patient results is 0.000 - 0.056 ng/ml.All other samples produced clinically concordant results when recalculated using a calibration from the correct calibrator.Based upon the instrument data, the effect of using the rxl calibrator with the loci method resulted in a depression of the result between 8 to 39 %.On 08/10/2016 the siemens technical applications specialist (tas) reviewed calibrations due to tni not auto accepting.The customer discovered that they had purchased the rxl ctni calibrator (rc421c) instead of loci tni (rc621).All calibrations that had not auto accepted had failed standard deviation (sd) for one or more levels but were accepted manually.The issue was resolved by recalibration of the ctni method with the appropriate calibrator set and customer education.The device is performing within specifications.No further evaluation of the device is required.
|
|
Event Description
|
Biased low troponin i (tni) results were obtained on qc and patient samples during a period from (b)(6) 2016 through (b)(6) 2016 after an incorrect calibrator set had been used to calibrate the tni method.Qc remained within laboratory ranges and patient results were reported to the physicians.After the calibration error was identified, patient and qc samples were recalculated and higher results were obtained.There is no indication that patient treatment was altered or prescribed on the basis of biased low troponin i results.There was no report of adverse health consequences as a result of the biased low troponin i results.
|
|
Search Alerts/Recalls
|