• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COVIDIEN KANGAROO JOEY,PUMP W/POLE CLMP; ENTERAL FEEDING PUMP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COVIDIEN KANGAROO JOEY,PUMP W/POLE CLMP; ENTERAL FEEDING PUMP Back to Search Results
Model Number 383400
Device Problems Excess Flow or Over-Infusion (1311); Improper Flow or Infusion (2954); Pumping Problem (3016)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
Submit date: 09/07/2016.An investigation is currently underway.Upon completion, the results will be forwarded.
 
Event Description
It was reported to covidien on (b)(6) 2016 that the customer experienced an issue with an enteral feeding pump.Customer reported the pump was not infusing correctly.On (b)(6) 2016, the customer stated the pump was set to infuse 900ml at 50ml per hour but 1000ml was delivered over 19 hours because the unit failed to suspend at the end of the set feeding.No harm to the patient or medical invention required.
 
Manufacturer Narrative
An evaluation of the kangaroo joey pump was performed and the customer states ¿the pump was set to infuse 900ml at 50ml per hour but 1000ml was delivered over 19 hours because the unit failed to suspend at the end of the set feeding.No harm to the patient or medical invention required¿.The unit was triaged and the reported issue could not be confirmed at this time.All device history records are reviewed for quality inspections and parameter compliance prior to releasing the product for shipment.Complaint trending information is being reviewed on a monthly basis and if a trend is observed, actions will be taken as necessary.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
KANGAROO JOEY,PUMP W/POLE CLMP
Type of Device
ENTERAL FEEDING PUMP
Manufacturer (Section D)
COVIDIEN
2824 airwest boulevard
plainfield IN 46168
Manufacturer (Section G)
COVIDIEN
2824 airwest boulevard
plainfield IN 46168
Manufacturer Contact
edward almeida
15 hampshire st
mansfield, MA 02048
5084524155
MDR Report Key5932960
MDR Text Key54093554
Report Number3008361498-2016-00076
Device Sequence Number1
Product Code LZH
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 08/21/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/07/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Model Number383400
Device Catalogue Number383400
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Date Manufacturer Received05/11/2016
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-