Catalog Number VLFT10GEN |
Device Problem
Failure to Obtain Sample (2533)
|
Patient Problem
Superficial (First Degree) Burn (2685)
|
Event Date 08/09/2016 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).To date the unit has not been received for evaluation.If the sample is received or if additional information pertinent to the incident is obtained a follow-up report will be submitted.A review of the device history records indicated that this serial number was released meeting all specifications as manufactured.
|
|
Event Description
|
The customer reported that during a mastectomy an arc was made between the handpiece being used with a ft10 generator and the port guide wire of a left subclavian bard port catheter resulting in a sponge and drapes adjacent to the field to ignite.The fire was stamped out in seconds and the field was covered with a wet sponge.Patient received a small 2x1 first degree burn to the skin.The area was dressed with polysporin and gauze.The customer has reported this incident under mw5064092.
|
|
Manufacturer Narrative
|
(b)(4).The unit was returned and the investigation did not identify anything that would have caused or contributed to the reported event.The unit was found to function normally and within specification.
|
|