COVIDIEN LP FT SERIES VESSEL SEALING GENERATOR; ELECRTOSURGICAL GENERATOR Back to Search Results
Catalog Number VLFT10GEN
Device Problem Failure to Obtain Sample (2533)
Patient Problem Superficial (First Degree) Burn (2685)
Event Date 08/09/2016
Event Type  malfunction  
Manufacturer Narrative
(b)(4).To date the unit has not been received for evaluation.If the sample is received or if additional information pertinent to the incident is obtained a follow-up report will be submitted.A review of the device history records indicated that this serial number was released meeting all specifications as manufactured.
 
Event Description
The customer reported that during a mastectomy an arc was made between the handpiece being used with a ft10 generator and the port guide wire of a left subclavian bard port catheter resulting in a sponge and drapes adjacent to the field to ignite.The fire was stamped out in seconds and the field was covered with a wet sponge.Patient received a small 2x1 first degree burn to the skin.The area was dressed with polysporin and gauze.The customer has reported this incident under mw5064092.
 
Manufacturer Narrative
(b)(4).The unit was returned and the investigation did not identify anything that would have caused or contributed to the reported event.The unit was found to function normally and within specification.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FT SERIES VESSEL SEALING GENERATOR
Type of Device
ELECRTOSURGICAL GENERATOR
Manufacturer (Section D)
COVIDIEN LP
5920 longbow drive
boulder CO 80301
Manufacturer Contact
sharon murphy
5920 longbow drive
boulder, CO 80301
2034925267
MDR Report Key5936738
MDR Text Key54276177
Report Number1717344-2016-00785
Device Sequence Number1
Product Code GEI
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K151649
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date 08/11/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/08/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberVLFT10GEN
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/23/2016
Is the Reporter a Health Professional? Yes
Date Manufacturer Received09/15/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/30/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-