Model Number TPO100B
Device Problem Device Alarm System (1012)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
Should additional information become available, a supplemental record will be filed.
 
Event Description
Provider states the unit isn't alarming and there isn't any error on the screen.
 
Manufacturer Narrative
Additional/updated information was added to reflect the device being returned to the manufacturer for evaluation.The result of the evaluation was that the pcb had an open circuit, causing the alarms not to function and no error on the screen, which confirmed the original complaint issue.
 
Event Description
Provider states the unit isn't alarming and there isn't any error on the screen.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SOLO2 TRANSPORTABLE OXYGEN CONCENTRATOR 9153650707
Type of Device
GENERATOR, OXYGEN, PORTABLE
Manufacturer (Section D)
INVACARE FLORIDA OPERATIONS
2101 east lake mary blvd
sanford FL 32773
Manufacturer (Section G)
INVACARE FLORIDA OPERATIONS
2101 east lake mary blvd
sanford FL 32773
Manufacturer Contact
kevin guyton
one invacare way
elyria, OH 44035
8003336900
MDR Report Key5939661
MDR Text Key54750698
Report Number1031452-2016-03498
Device Sequence Number1
Product Code CAW
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor
Reporter Occupation Medical Equipment Company Technician/Representative
Type of Report Initial,Followup
Report Date 09/23/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberTPO100B
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/23/2016
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 08/15/2016
Initial Date FDA Received09/09/2016
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received10/11/2016
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-