Results: the indigo system aspiration catheter 6 (cat6) was kinked in multiple locations and ovalized from approximately 14.0 cm from the distal tip to the distal tip.Conclusions: evaluation of the returned device revealed the cat6 was kinked in multiple locations and ovalized in the distal shaft.This damage may have occurred due to forceful manipulation of the cat6 during insertion or advancement into a sheath.Penumbra catheters are inspected during in-process inspection and during quality inspection after manufacturing.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.
|
The patient was undergoing a thrombectomy procedure to treat a chronic total occlusion (cto) using an indigo system aspiration catheter 6 (cat6).During the procedure, the physician advanced the cat6 into another manufacturer's sheath but noticed that distal end of the cat6 was crushed.Therefore, the cat6 was removed and the procedure was completed using a new cat6.It should be noted that other manufacturer's cto crossing device and atherectomy devices were also used during the procedure.There was no report of an adverse effect on the patient.
|