MAUDE Adverse Event Report: SYNCARDIA SYSTEMS, INC. SYNCARDIA FREEDOM DRIVER; EXTERNAL PNEUMATIC DRIVER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SYNCARDIA SYSTEMS, INC. SYNCARDIA FREEDOM DRIVER; EXTERNAL PNEUMATIC DRIVER Back to Search Results
Catalog Number 595000-001
Device Problem Device Displays Incorrect Message (2591)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 08/18/2016
Event Type  malfunction  
Manufacturer Narrative
(b)(4) initial.
 
Event Description
The customer reported that the freedom driver exhibited a fault alarm while supporting a patient sitting in his hospital room watching tv.The customer also reported that the patient was subsequently switched to the backup freedom driver.There was no reported adverse patient impact.The freedom driver was returned to syncardia for evaluation.Visual inspection of the external and internal components revealed no abnormalities.The driver was tested and passed all pressure test requirements, which included cardiac output, right atrial pressure (rap), aortic pressure (aop), pulmonary arterial pressure (pap) and left atrial pressure (lap) performance metrics associated with nominal normotensive and hypertensive patient simulator settings.Review of the alarm history (eeprom) revealed a fault alarm code that is indicative of a potentially failing u22 pressure sensor on the main pcba (printed circuit board assembly) and is consistent with the alarm that the customer experienced.Further investigation of the main pcba confirmed that the root cause for the reported fault alarm was a failure of the pressure sensor (u22) on the main pcba.The driver was serviced, including replacement of the main pcba, and passed all final performance testing.Syncardia initiated a corrective action (capa) to prevent u22 failures, which is in the process of being implemented.This failure mode posed a low risk to the patient because it did not prevent the freedom driver from performing its life-sustaining functions.This issue will continue to be monitored and trended as part of the customer experience process.Syncardia has completed its evaluation of this complaint and is closing this file.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNCARDIA FREEDOM DRIVER
Type of Device
EXTERNAL PNEUMATIC DRIVER
Manufacturer (Section D)
SYNCARDIA SYSTEMS, INC.
1992 e. silverlake road
tucson AZ 85713
Manufacturer Contact
carole marcot
1992 e. silverlake road
tucson, AZ 85713
5205451234
MDR Report Key5943120
MDR Text Key54444309
Report Number3003761017-2016-00305
Device Sequence Number1
Product Code LOZ
UDI-Device Identifier00858000003121
UDI-Public(01)00858000003121
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P030011
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Reporter Occupation Other Health Care Professional
Remedial Action Replace
Type of Report Initial
Report Date 08/18/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/12/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number595000-001
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/29/2016
Is the Reporter a Health Professional? Yes
Date Manufacturer Received08/19/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/02/2012
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Age55 YR
-
-