Catalog Number 400207 |
Device Problem
Noise, Audible (3273)
|
Patient Problems
No Consequences Or Impact To Patient (2199); No Known Impact Or Consequence To Patient (2692)
|
Event Date 09/02/2016 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).
|
|
Event Description
|
The customer reported that there was a loud humming noise coming from the css console while it was supporting a patient.The customer also reported that the patient was switched to a backup driver.There was no reported adverse patient impact.This alleged failure mode poses a low risk to the patient because it did not prevent the css console from performing its life-sustaining functions.The css console will be returned to syncardia for evaluation.The results of the investigation will be provided in a follow-up mdr.
|
|
Manufacturer Narrative
|
The css console was returned to syncardia for evaluation.During investigation testing, the customer-reported humming noise was verified.The root cause was determined to be a malfunction of the power supply.This issue will continue to be monitored and trended as part of the customer experience process.Syncardia has completed its evaluation of this complaint and is closing this file.(b)(4) follow-up report 1.
|
|
Event Description
|
The customer reported that there was a loud humming noise coming from the css console while it was supporting a patient.The customer also reported that the patient was switched to a backup driver.There was no reported adverse patient impact.
|
|