• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COVIDIEN LP FORCETRIAD ENERGY PLATFORM; ELECTROSURGICAL GENERATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COVIDIEN LP FORCETRIAD ENERGY PLATFORM; ELECTROSURGICAL GENERATOR Back to Search Results
Catalog Number FORCETRIAD
Device Problem Failure to Obtain Sample (2533)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Date of initial report : (b)(6) 2016.To date the unit has not been received for evaluation.If the sample is received or if additional information pertinent to the incident is obtained a follow-up report will be submitted.
 
Event Description
The customer reported that prior to a procured and during testing an unknown type of monopolar pencil could be activated without a rem plate while used with a forcetriad generator.
 
Manufacturer Narrative
(b)(4).Date of follow-up report: 10/25/2016.The reported condition was confirmed.The investigation found the generator powered up with the rem icon illuminating green indicating that the generator is capable of delivering energy output.The investigation isolated the failure to the rem calibration, but a root cause was not identified.The probable root cause could not be determined based on the information provided.A complete calibration was performed on the generator to address the condition.No trend has been identified.No corrective action is required, because no trend has been identified.A review of the appropriate device history records indicates that this unit was upgraded and was released meeting all specifications.The appropriate upgrades were implemented.The returned product was calibrated and testing found the generator functions normally.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FORCETRIAD ENERGY PLATFORM
Type of Device
ELECTROSURGICAL GENERATOR
Manufacturer (Section D)
COVIDIEN LP
5920 longbow drive
boulder CO 80301
Manufacturer Contact
sharon murphy
5920 longbow drive
boulder, CO 80301
2034925267
MDR Report Key5945504
MDR Text Key54525117
Report Number1717344-2016-00843
Device Sequence Number1
Product Code GEI
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K110268
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date 08/17/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/13/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberFORCETRIAD
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/18/2016
Is the Reporter a Health Professional? Yes
Date Manufacturer Received10/24/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/31/2009
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Treatment
UNKNOWN MONOPOLAR PENCIL
-
-