Catalog Number FORCETRIAD |
Device Problem
Failure to Obtain Sample (2533)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).Date of initial report : (b)(6) 2016.To date the unit has not been received for evaluation.If the sample is received or if additional information pertinent to the incident is obtained a follow-up report will be submitted.
|
|
Event Description
|
The customer reported that prior to a procured and during testing an unknown type of monopolar pencil could be activated without a rem plate while used with a forcetriad generator.
|
|
Manufacturer Narrative
|
(b)(4).Date of follow-up report: 10/25/2016.The reported condition was confirmed.The investigation found the generator powered up with the rem icon illuminating green indicating that the generator is capable of delivering energy output.The investigation isolated the failure to the rem calibration, but a root cause was not identified.The probable root cause could not be determined based on the information provided.A complete calibration was performed on the generator to address the condition.No trend has been identified.No corrective action is required, because no trend has been identified.A review of the appropriate device history records indicates that this unit was upgraded and was released meeting all specifications.The appropriate upgrades were implemented.The returned product was calibrated and testing found the generator functions normally.
|
|