PRODUCTOS PARA EL CUIDADO DE LA SALUD -9611590 BARD DISPOZ-A-BAG LEG BAG WITH FLIP-FLO DRAINAGE VALVE, FABRIC LEG Back to Search Results
Catalog Number 150819
Device Problem Restricted Flow rate (1248)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Type  malfunction  
Manufacturer Narrative
The device was not returned for evaluation.The lot number is unknown; therefore, the device history record could not be reviewed.The instructions for use states the following: "- directions for use: separate notches within circles.Pull straps through holes and around leg.Position bag on leg with flutter valve at top.Attach catheter or extension tubing to top inlet.When wearing bag below knee, attach bard extension tubing (catalog no.150615 or 4a4194).When using extension tubing, make sure to connect tube at least to the 3rd taper of the connector.To empty dispoz-a-bag, push green lever on flip-flo valve out and down.Important: be sure to reclose flip-flo valve after emptying bag." (b)(4).The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.The device was not returned.
 
Event Description
It was reported that there were issues with the urine emptying into the leg bag.It was later reported that the urine was not draining into the bag at all; it is remaining in the top portion of the tubing.The complainant stated that when disconnecting the tubing from the leg bag, it was like "a vacuum is released and the urine drains out quickly".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BARD DISPOZ-A-BAG LEG BAG WITH FLIP-FLO DRAINAGE VALVE, FABRIC LEG
Type of Device
LEG BAG
Manufacturer (Section D)
PRODUCTOS PARA EL CUIDADO DE LA SALUD -9611590
km. 7 carretera internacional
nogales, sonora 85621
MX  85621
Manufacturer (Section G)
PRODUCTOS PARA EL CUIDADO DE LA SALUD -9611590
km. 7 carretera internacional
nogales, sonora 85621
MX   85621
Manufacturer Contact
janna parks
8195 industrial blvd
covington, GA 30014
7707846100
MDR Report Key5956261
MDR Text Key55263963
Report Number1018233-2016-01235
Device Sequence Number1
Product Code FAQ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,other,use
Reporter Occupation Nurse
Type of Report Initial
Report Date 08/24/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number150819
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 08/24/2016
Initial Date FDA Received09/16/2016
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-