Device Problem Tidal Volume Fluctuations (1634)
Patient Problem No Patient Involvement (2645)
Event Date 08/17/2016
Event Type  malfunction  
Manufacturer Narrative
No report of patient involvement.The initial reporter is located outside the u.S., and therefore this information is not provided due to country privacy laws.A 3rd party service representative performed a checkout of the equipment and confirmed the reported complaint.The inspiratory valve calibration was performed, and the unit was returned to service.
 
Event Description
The hospital reported the unit alarmed and was unable to ventilate adequately in volume mode.There was no report of patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AESPIRE 7100
Type of Device
ANESTHESIA GAS MACHINE
Manufacturer (Section D)
DATEX-OHMEDA, INC.
3030 ohmeda drive
madison WI 53718
Manufacturer (Section G)
DATEX-OHMEDA, INC.
3030 ohmeda drive
madison WI 53718
Manufacturer Contact
joseph seliga
3000 n grandview blvd.
waukesha, WI 53188
MDR Report Key5956484
MDR Text Key54938715
Report Number2112667-2016-01815
Device Sequence Number1
Product Code BSZ
Combination Product (y/n)N
PMA/PMN Number
K000706
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,u
Type of Report Initial
Report Date 09/16/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 08/17/2016
Initial Date FDA Received09/16/2016
Date Device Manufactured06/12/2006
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-