Suspect lot review: a review of suspect lot#2892210 showed that (b)(4) units were manufactured, tested, inspected and released in july 2014, citing no anomalies.Receiving inspection: 09/06/2016 - received one new a1067, 9" smallbore quadfuse ext.Set.Lot# 2892210 and one used a1067, 9" smallbore quadfuse ext.Set.Unknown.No mating devices were returned.The device was not decontaminated.Functional testing:one new and one used a1067 quadfuse extension sets were returned for investigation of leakage.The two a1067 quadfuse extension sets were pressure leak tested.The used a1067 had leakage at two of the bonds between the wingless female luer and the male luer of the nanoclave.There was no leakage at any location along the fluid path of the new a1067 quadfuse extension set.Final analysis summary: the complaint of a1067 quadfuse leakage was confirmed with the used set.The leakage was the result of channel leakage between the wingless female luer and the male luer of the nanoclave due to an insufficient bond.The new a1067 quadfuse did not leak at any location along the fluid path.Additional investigation activities and engineering efforts are in progress to improve the bonding connection.As an interim measure heightened awareness have been initiated.
|
Complaint received regarding three a1067, 9" smallbore quadfuse ext.Set, lot number unknown.Leaking noted around the nanoclave ports while infusing dopamine, occurred 3 times on one patient.Noticed as b/p failed to respond and fluid noticed to be leaking at the nanoclave.This was a problem in (b)(6) lot number changed but the a1063 is a new problem.Delay in critical therapy reported.No adverse patient consequences reported.
|