A peritoneal dialysis patient reported that she had just finished draining 8,459ml and expected 2,000ml in drain 0.Patient reported feeling bloated but no pain.Additional medical records have been requested.The reported large drain of 8,459ml was 423% over the expected drain volume which resulted in a reportable device malfunction.
|
Additional information: an external, visual inspection of the returned cycler exterior showed no signs of any physical damage.The heater tray/scale was not obstructed.Two simulated treatments were performed and completed without any alarms or problems occurring.Neither reported complaint system could be reproduced during testing.A simulated treatment was performed in which the amount of fluid delivered to a pseudo-patient bag during each fill phase was weighed.The weighed fluid volumes were compared against the programmed fill value, and the weighed volumes for each fill phase were determined to be within expected tolerance for the liberty cycler.The simulated treatment was performed and completed without any problems occurring.During the treatment test, a groaning or burping sound was observed during pump operation of the cycler.An accelerated stress test identified a weak pump a motor.Grinding noise due to the weak motors was observed during the test.The valve actuation test passed.The system air leak test passed.The patient sensor calibration check passed.The load cell value and verification was within tolerance.There were no internal, visual discrepancies found in the inspection of the received cycler unit.An investigation of the device manufacturing records was conducted by the manufacturer.There were no deviations or non-conformances during the manufacturing process.In addition, the device history record review confirmed the labeling, material, and process controls were within specification.
|