• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MDT SOFAMOR DANEK PUERTO RICO MFG CD HORIZON SPINAL SYSTEM; ORTHOSIS, SPINAL PEDICLE FIXATION, FOR DEGENERATIVE DISC DISEASE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MDT SOFAMOR DANEK PUERTO RICO MFG CD HORIZON SPINAL SYSTEM; ORTHOSIS, SPINAL PEDICLE FIXATION, FOR DEGENERATIVE DISC DISEASE Back to Search Results
Catalog Number 7078396
Device Problems Metal Shedding Debris (1804); Failure to Align (2522)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 08/23/2016
Event Type  malfunction  
Manufacturer Narrative
Product was not returned to the manufacturer for evaluation.Therefore, we are unable to determine the definitive cause of the r eported event.
 
Event Description
It was reported that the patient underwent a posterior correction fusion surgery at t10-s2ai levels, for the treatment of spinal canal stenosis.During surgery, when the surgeon tried to break off a set screw for final tightening using a break off driver after he picked up mas 8.5-70 which was inserted at left of s2ai using a lateral connector, but he could not break off the set screw.And burrs occurred because the set screw kept turning.The surgeon changed the set screw to new one but the same thing happened in the try.The surgeon stopped the tightening before a set screw is spun and the surgery was finished without break off.No patient complications were reported.
 
Manufacturer Narrative
Product analysis :visual and microscopic examination of the break-off set screws identified localized thread damage and shearing consistent with overloading the set screw during attempted construct assembly.The nature of overload condition suggests the rod may not have been fully seated in the mas saddle, thus placing rod reduction load onto set screw threads during attempted assembly.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Manufacturer Narrative
A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CD HORIZON SPINAL SYSTEM
Type of Device
ORTHOSIS, SPINAL PEDICLE FIXATION, FOR DEGENERATIVE DISC DISEASE
Manufacturer (Section D)
MDT SOFAMOR DANEK PUERTO RICO MFG
barrio marianna rd 909, km0.4
humacao PR 00792
Manufacturer (Section G)
MDT SOFAMOR DANEK PUERTO RICO MFG
barrio marianna rd 909, km0.4
humacao PR 00792
Manufacturer Contact
greg anglin
1800 pyramid place
memphis, TN 38132
9013963133
MDR Report Key5965947
MDR Text Key55212878
Report Number1030489-2016-02618
Device Sequence Number1
Product Code NKB
Combination Product (y/n)N
Reporter Country CodeJA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date 10/19/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/21/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date11/25/2023
Device Catalogue Number7078396
Device Lot NumberH5235356
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/13/2016
Is the Reporter a Health Professional? Yes
Date Manufacturer Received10/19/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/25/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-