Catalog Number 900-600
Device Problems Bent (1059); Crack (1135)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 06/17/2015
Event Type  malfunction  
Manufacturer Narrative
The reported defective device has yet to be returned to the manufacturer for a device evaluation.The firm is attempting to obtain the device.An investigation into the root cause for product problem is currently in progress.The results of the device evaluation will be sent via a follow up medwatch.Complaint # (b)(4).Device has not been returned to date.
 
Event Description
As reported (b)(6) 2016, a patient of unknown age and gender presented for a microwave procedure of the liver.It was reported during the procedure, the applicator tip was bending.The tip remained fully attached to the applicator shaft while in the patient.The applicator was set aside and a new of the same device was used to successfully complete the procedure.It was reported the patient did not suffer harm or injury due to the event.It was reported the disposable device is available for return to the manufacturer for evaluation.
 
Manufacturer Narrative
The customers reported complaint of a broken fiber could not be confirmed because no sample was returned for evaluation.Without confirming the complaint, a root cause cannot be determined.The patient was unaffected due to this event.A review of the lot history records was performed for the reported packaging lot for any deviations related to the reported defect of the complaint.The review confirms that the lots met all material, assembly, and performance specifications.At the vendor facility, during the manufacturing process, fibers are repeatedly bent and coiled and there is a final visual inspection performed with a hene laser to check the entire fiber length for defects prior to shipment.During the packaging step, the fibers are inspected for damage.A review of the device history records was performed for the reported packaging and component lots for any deviations related to the reported defect of the complaint.The review confirms that the lots met all material, assembly, and performance specifications.The instructions for use, which is supplied to the end user with this catalog number contains the following statement "prior to and during use, avoid damaging the fiber by striking, stressing or excessive bending.Do not coil the fiber tighter than a radius of 60 mm".A review of the angiodynamics complaint system noted no adverse trends for this complaint type and product family.This type of complaint will continue to be monitored for trends.(b)(4).Device not available for return.
 
Manufacturer Narrative
Returned for evaluation was one pmta accu2i standard applicator.A visual examination of the device noted that the tip of the applicator was broken.Functional testing could not be performed due to the condition of the returned device.The customer's reported complaint description of the tip fracture is confirmed.A review of the device history records was performed for the reported packaging and component lots for any deviations related to the reported defect of the complaint.The review confirms that the lots met all material, assembly, and performance specifications.Based on the device history review this event does not appear to be due to a manufacturing defect.A possible contributing factor could have been operational context; i.E.Probe placement into hard tissue or lateral forces on the applicator tip during placement or removal.The ifu states; "avoid placing lateral forces on the applicator tip during placement or removal and lateral forces on the applicator tip should be avoided both during insertion and removal.Failure to do so could result in damage to the microwave array, failure of the applicator and injury to the patient." a review of the angiodynamics complaint system noted no trends for this complaint type and product family.This type of complaint will continue to be monitored for trends.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PMTA ACCU2I STANDARD APPLICATOR
Type of Device
MICROWAVE APPLICATOR
Manufacturer (Section D)
ANGIODYNAMICS
603 queensbury avenue
queensbury NY 12804
Manufacturer (Section G)
ANGIODYNAMICS, INC.
603 queensbury avenue
queensbury NY 12804
Manufacturer Contact
law ryan
603 queensbury avenue
queensbury, NY 12804
5187424488
MDR Report Key5966068
MDR Text Key55224208
Report Number1319211-2016-00136
Device Sequence Number1
Product Code NEY
Combination Product (y/n)N
Reporter Country CodeTU
PMA/PMN Number
K122762
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,foreign
Reporter Occupation Medical Equipment Company Technician/Representative
Type of Report Initial,Followup,Followup
Report Date 09/01/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Physician
Device Expiration Date05/31/2018
Device Catalogue Number900-600
Device Lot Number15060348
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/14/2016
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 09/01/2016
Initial Date FDA Received09/21/2016
Supplement Dates Manufacturer ReceivedNot provided
Not provided
Supplement Dates FDA Received01/09/2017
01/10/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/20/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-