U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ROCHE DIAGNOSTICS FT4, FREE THYROXINE; RADIOIMMUNOASSAY, FREE THYROXINE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ROCHE DIAGNOSTICS FT4, FREE THYROXINE; RADIOIMMUNOASSAY, FREE THYROXINE Back to Search Results
Catalog Number 06437281190
Device Problem Incorrect Or Inadequate Test Results (2456)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 09/06/2016
Event Type  malfunction  
Manufacturer Narrative
This event occurred in (b)(6).
 
Event Description
The customer questioned results from 1 patient sample tested for thyrotropin (tsh), elecsys ft4 ii assay (ft4 ii) and elecsys ft3 iii (ft3 iii).The customer provided the sample for investigation.Of the data provided, erroneous ft4 ii and ft3 iii results were identified between the customer's e602 analyzer and an e411 analyzer used at the investigation site.The customer reported the results from the e602 analyzer outside of the laboratory.This medwatch will cover ft4 ii.Refer to medwatch with patient identifier (b)(6) for information on the ft3 iii erroneous results.No adverse event occurred.The e411 analyzer serial number was (b)(4).The ft4 ii reagent lot number used at the investigation site with the e411 analyzer was 168540 with an expiration date of 07/31/2017.The serial number for the customer's e602 analyzer is not known.
 
Manufacturer Narrative
Upon further investigation of the patient sample, a streptavidin interference was confirmed.This most likely caused the high ft4 ii and ft3 iii results.The reagent used to perform the tests contains streptavidin.This specific interference is addressed in product labeling.For diagnostic purposes, the results should always be assessed in conjunction with the patient¿s medical history, clinical examination and other findings.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FT4, FREE THYROXINE
Type of Device
RADIOIMMUNOASSAY, FREE THYROXINE
Manufacturer (Section D)
ROCHE DIAGNOSTICS
9115 hague road
indianapolis IN 46250 0457
Manufacturer (Section G)
ROCHE DIAGNOSTICS GMBH
sandhoferstrasse 116
na
mannheim (baden-wurttemberg) 68305
GM   68305
Manufacturer Contact
michael leslie
9115 hague road
na
indianapolis, IN 46250
3175214343
MDR Report Key5967713
MDR Text Key55319088
Report Number1823260-2016-01428
Device Sequence Number1
Product Code CEC
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K961489
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,u
Reporter Occupation Health Professional
Type of Report Initial,Followup
Report Date 09/28/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/21/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number06437281190
Device Lot NumberASKU
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received09/07/2016
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-