(b)(4).The hospital did not provide an rt202 breathing circuit or mr290 chamber for evaluation as they were not reporting any defect of the devices.Rather, this was a general complaint concerning the circuit disconnecting from the chamber when put under excessive strain as a result of pulling on the system or having it caught on a bed that is being raised.Based on the description of events provided by the hospital, we can conclude that the rt202 circuit and the mr290 chamber were operating correctly but had disconnected when under excessive pulling force.When properly connected the circuit and chamber form a tight fit and cannot disconnect spontaneously.The connection between the circuit elbow and chamber port is a standard medical taper fit and conforms to iso (b)(4): anaesthetic and respiratory equipment -- conical connectors.All breathing circuits chambers are pressure tested for leaks during production and those that fail are rejected.The user instructions that accompany the rt202 adult breathing circuit contain the following: check all connections are tight before use.Set appropriate ventilator alarms.Do not stretch or milk the tubing.
|