DENTSPLY PROFESSIONAL ESTYLUS 1:5 HIGH SPEED CONTRA ANGLE ATTACHMENT; HANDPIECE, AIR-POWERED, DENTAL Back to Search Results
Catalog Number 774105
Device Problem Overheating of Device (1437)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Type  malfunction  
Manufacturer Narrative
Maximum temperature recorded on the head of the attachment during free run testing did exceed the maximum allowable temperature.The attachment stopped working after 1 minute and 30 seconds of free run testing.Cap end bearing failure, free roaming ball bearings and excessive debris most likely created friction within the head which resulted in temperature increase.It is reasonable to suspect gear function was compromised by the added debris inside the head which is evident from significant, abrasive wear on the teeth of the gears.All components looked dry with no evidence of lubrication.
 
Event Description
In this event a doctor reported that an estylus 1:5 ca overheated.There was no injury or intervention.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ESTYLUS 1:5 HIGH SPEED CONTRA ANGLE ATTACHMENT
Type of Device
HANDPIECE, AIR-POWERED, DENTAL
Manufacturer (Section D)
DENTSPLY PROFESSIONAL
901 west oakton st.
des plaines IL 60018
Manufacturer (Section G)
DENTSPLY PROFESSIONAL
901 west oakton st.
des plaines IL 60018
Manufacturer Contact
helen lewis
221 w. philadelphia st.
suite 60w
york, PA 17401
7178494229
MDR Report Key5974890
MDR Text Key55562417
Report Number1419322-2016-00246
Device Sequence Number1
Product Code EFB
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K031145
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type other
Reporter Occupation Service Personnel
Type of Report Initial
Report Date 08/25/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number774105
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/26/2016
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 08/25/2016
Initial Date FDA Received09/23/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/24/2011
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-