• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC - COSTA RICA (HEREDIA) ROTAWIRE¿ AND WIRECLIP¿ TORQUER; CATHETER, CORONARY, ATHERECTOMY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC - COSTA RICA (HEREDIA) ROTAWIRE¿ AND WIRECLIP¿ TORQUER; CATHETER, CORONARY, ATHERECTOMY Back to Search Results
Model Number H802228240022
Device Problems Kinked (1339); Device Damaged Prior to Use (2284)
Patient Problem No Patient Involvement (2645)
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Age at time of event: 18 years or older.(b)(4).Device evaluated by mfr: the device was returned for evaluation.Device analysis noted that the guidewire was broken at 207 cm from the distal section.The spring tip was not returned.The overall length and the outer diameter (od) of proximal section were not measured due to the device condition (wire broken in the middle of the device).The od of the distal tip and the middle of the device near the broke section is within specification.The manufacturing batch record review confirmed that the device met all material, assembly and performance specifications.The root cause is not confirmed as there was no evidence of the alleged issue or any anomalies which could have contributed to the reported difficulty.(b)(4).
 
Event Description
Reportable based on device analysis completed on 30aug2016.It was reported that a guidewire kink occurred.The 90% stenosed target lesion was located in the moderately tortuous and severely calcified left anterior descending (lad) artery.A rotawire¿ and wireclip¿ torquer was selected for use.Upon preparation, outside the patient's body, it was noticed that the rotawire was kinked.The procedure was completed with another of the same device.No patient complications were reported and the patient's status is good.However, device analysis revealed the wire was broken in the middle of the device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ROTAWIRE¿ AND WIRECLIP¿ TORQUER
Type of Device
CATHETER, CORONARY, ATHERECTOMY
Manufacturer (Section D)
BOSTON SCIENTIFIC - COSTA RICA (HEREDIA)
302 parkway
la aurora
heredia
CS 
Manufacturer (Section G)
BOSTON SCIENTIFIC - COSTA RICA (HEREDIA)
302 parkway
la aurora
heredia
CS  
Manufacturer Contact
linda leimer
one scimed place
maple grove, MN 55311
7634941700
MDR Report Key5977960
MDR Text Key55691927
Report Number2134265-2016-08342
Device Sequence Number1
Product Code MCX
Combination Product (y/n)N
PMA/PMN Number
P900056
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,distri
Reporter Occupation Physician
Type of Report Initial
Report Date 08/30/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date03/15/2018
Device Model NumberH802228240022
Device Catalogue Number22824-002
Device Lot Number19021361
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/10/2016
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 08/30/2016
Initial Date FDA Received09/26/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/17/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-