• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SYNTHES MONUMENT DEPTH GAUGE FOR 2.0MM AND 2.4MM SCREWS; GAUGE, DEPTH

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SYNTHES MONUMENT DEPTH GAUGE FOR 2.0MM AND 2.4MM SCREWS; GAUGE, DEPTH Back to Search Results
Catalog Number 319.006
Device Problem Break (1069)
Patient Problems No Consequences Or Impact To Patient (2199); No Patient Involvement (2645)
Event Type  malfunction  
Manufacturer Narrative
No patient involvement reported.Event: unknown when device broke.Device is an instrument and is not implanted/explanted.Date returned to manufacturer subject device has been received and is currently in the evaluation process.Investigation is ongoing; no conclusion could be drawn as product is entering the complaint system.Device history records review was conducted.Dhr review for: manufacturing location: (b)(4), manufacturing date: 16-mar-2011, part #: 319.006, lot#: 6617554 (non-sterile) - depth gauge for 2.0mm and 2.4mm screws.Qy.16.No ncrs were generated during production.Review of the device history record(s) showed that there were no issues during the manufacture of the product that would contribute to this complaint condition.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
It was reported on (b)(6) 2016, that the depth gauge for 2.0mm and 2.4mm screws was found broken in a box in trunk of car while moving.There was no patient involvement and there is no additional information to provide.This complaint involves 1 device.This report is 1 of 1 for (b)(4).
 
Manufacturer Narrative
A product investigation was completed: one depth gauge for 2.0mm and 2.4mm screws (part 319.006, lot 6617554) was returned for investigation.It was reported to have a broken tip/needle, and it is confirmed since the device was received in multiple pieces.The hooked needle stem of the device is broken off at the base of the black body (the broken stem is approximately 75mm in length) and was not returned.The slider is loose in the hollow body.The handle has various marks and scratches, and the laser marking on the shaft is clearly visible.The exact cause of the complaint condition cannot be determined.But the condition of the depth gauge is consistent with the result of a bending force applied to the needle portion of the depth gauge that is beyond the yield limit of the material.There is a protection sleeve to protect the needle during transport and the body that slides on the measuring portion adds additional protection to the needle attachment point during storage.This particular depth gauge is part of at least 14 technique guides, including the 2.4 mm variable angle lcp distal radius system and is used to measure the depth of the holes for the 2.0mm/2.4mm screws to ensure the correct screw length is used during the procedure.The relevant drawings for the device(s) were reviewed.No drawing issues or discrepancies were noted.The design is adequate for its intended use and did not contribute to this complaint condition.The thickness of the needle (1.25 mm) is driven by the fact that the needle must fit into a drilled hole of 1.5 mm, and the length (80 mm) is determined so the slider can measure screws up to 40 mm.The material of the needle probe component is extra hard 316ss, which is an appropriate material for an instrument component of this type.The design, materials and finishing processes were found to be appropriate for the intended use of these devices.A review of the device history records showed that there were no issues during the manufacture of the product that would contribute to this complaint condition.Although the exact cause cannot be determined, the most probable root cause for this complaint is excessive force exerted on the depth gauge by placing / dropping heavy instruments on top of the device during the sterilization process.During the investigation no product design issues or discrepancies were observed that may have contributed to the complaint condition.A visual inspection, complaint history review, drawing review, and risk assessment review were performed as part of this investigation.No product design issues or discrepancies were observed.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Manufacturer Narrative
A product investigation was completed: one depth gauge for 2.0mm and 2.4mm screws (part 319.006, lot 6617554) was returned for investigation.It was reported to have broken, and it is confirmed since the device was received in multiple pieces.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DEPTH GAUGE FOR 2.0MM AND 2.4MM SCREWS
Type of Device
GAUGE, DEPTH
Manufacturer (Section D)
SYNTHES MONUMENT
1051 synthes ave
monument CO 80132
Manufacturer (Section G)
SYNTHES MONUMENT
1051 synthes ave
monument CO 80132
Manufacturer Contact
mark vornheder
1302 wrights lane east
west chester, PA 19380
6107195000
MDR Report Key5978613
MDR Text Key55694361
Report Number1719045-2016-10709
Device Sequence Number1
Product Code HTJ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date 09/08/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/26/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number319.006
Device Lot Number6617554
Other Device ID Number(01)10886982189943(10)6617554
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/15/2016
Is the Reporter a Health Professional? No
Date Manufacturer Received11/04/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/16/2011
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-