Model Number 1095059
Device Problem Electrical /Electronic Property Problem (1198)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 09/06/2016
Event Type  malfunction  
Event Description
Our customer reported that the prong on the plug of this device was gone.On return of the device for evaluation, it was noted that one prong was missing from the plug.Our evaluation concluded that the plug may have been damaged by the end user applying excessive force, but there is a possibility that it broke off in the mains socket - if this was the case then a risk of exposure to a live part would be created.There is no report of any patient harm, and the device is not life-supporting or life-sustaining.The ifu for the device contains the warning 'the power supply cord cannot be replaced by the user.In case the power supply cord becomes damaged, contact philips respironics customer service for replacement (there is no service option)'.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INNOSPIRE ELEGANCE
Type of Device
COMPRESSOR, AIR, PORTABLE
Manufacturer (Section D)
RESPIRONICS RESPIRATORY DRUG DELIVERY (UK) LTD
chichester business park
city fields way, tangmere
chichester, west sussex PO20 2FT
UK  PO20 2FT
Manufacturer Contact
alessandro agosti
chichester business park
city fields way, tangmere
chichester, west sussex PO20 -2FT
UK   PO20 2FT
704231549
MDR Report Key5979294
MDR Text Key55669827
Report Number9681154-2016-00014
Device Sequence Number1
Product Code CAF
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K042655
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor
Reporter Occupation Other
Type of Report Initial
Report Date 09/06/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/27/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model Number1095059
Device Catalogue Number1100134
Device Lot Number150108
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/22/2016
Date Manufacturer Received09/06/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/08/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse