The customer stated that they received erroneous results for two samples from the same patient tested for roche diagnostics cobas elecsys anti-tpo (anti-tpo) and the elecsys tsh assay (tsh) on an e411 analyzer.Of the two samples, one had erroneous results that were reported outside of the laboratory for anti-tpo and tsh.This medwatch will refer to anti-tpo.Please refer to the medwatch with (b)(6) for information related to tsh.The sample initially resulted as 2.19 miu/l for tsh and 47.73 iu/ml for anti-tpo when tested on the e411 analyzer.The sample was tested on a siemens immulite analyzer, resulting as 1.635 uiu/ml for tsh and < 10 iu/ml for anti-tpo (normal reference range is less than 35 iu/ml).The results were released to the patient along with the suggestion to perform a new measurement in another laboratory.The patient was not adversely affected.The e411 analyzer serial number was (b)(4).During investigations, it was stated that the anti-tpo cut-off value is higher than the normal cut-off, so the anti-tpo result from (b)(6) 2016 can be considered to be negative.The absolute values of measurement results achieved with assays from different vendors cannot always be directly compared as each assay has different standardization procedures and different setups (antibodies used and other components).
|