STRYKER ORTHOPAEDICS-MAHWAH TRIATHLON CR FEM COMP #7 R-CEM; PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, CO Back to Search Results
Catalog Number 5510F702
Device Problems Difficult to Remove (1528); Manufacturing, Packaging or Shipping Problem (2975); Packaging Problem (3007)
Patient Problems No Consequences Or Impact To Patient (2199); No Information (3190)
Event Date 08/30/2016
Event Type  malfunction  
Manufacturer Narrative
Additional information has been requested.Should additional information become available it will be reported in a supplemental report upon completion of the investigation.
 
Event Description
Sales rep informed by phone the following: nurse (b)(6) had called and informed that they couldn't get the item ref 5510f702 out of the box in a sterile manner.Customer took another item and operation was completed successfully.There was a short delay during the operation when a new item was taken to continue the operation.
 
Manufacturer Narrative
An event regarding packaging damage involving a triathlon femoral component was reported.The event was confirmed.Method & results: -device evaluation and results: based on the visual inspection of the returned packaging it appears that this component packaging was subjected to excessive/inappropriate handling whereby the carton may have been compressed and/or dropped from a height causing damage to the carton making it difficult to remove the device.-medical records received and evaluation: not performed as the event is related to a packaging issue and no adverse consequences to the patient were reported.-device history review: dhr review for the reported lot determined that the device was manufactured and packed to specification.-complaint history review: there have been no other similar events for the lot referenced.Conclusions: the investigation concluded that the packaging damage was most likely caused by excessive/incorrect handling prior to the opening packaging whereby the carton may have been compressed and/or dropped from a height causing the damage to the carton making it difficult to remove the device.No further investigation for this event is required at this time.
 
Event Description
Sales rep informed by phone the following: nurse sk had called and informed that they couldn't get the item ref 5510f702 out of the box in a sterile manner.Customer took another item and operation was completed successfully.There was a short delay during the operation when a new item was taken to continue the operation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRIATHLON CR FEM COMP #7 R-CEM
Type of Device
PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, CO
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer (Section G)
STRYKER ORTHOPAEDICS-LIMERICK
raheen business park
limerick NA
Manufacturer Contact
keyla colon
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key5980529
MDR Text Key56022070
Report Number0002249697-2016-03050
Device Sequence Number1
Product Code MBH
Combination Product (y/n)N
Reporter Country CodeFI
PMA/PMN Number
K141056
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,other
Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date 08/30/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date05/31/2021
Device Catalogue Number5510F702
Device Lot NumberA3C2H
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/19/2016
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 08/30/2016
Initial Date FDA Received09/27/2016
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received12/05/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/09/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
-
-