DEPUY SYNTHES POWER TOOLS SMALL BATTERY DRIVE; INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED, MOTOR/ACCESS AND ATTACHMENT
|
Back to Search Results |
|
Catalog Number 532.010 |
Device Problems
Sticking (1597); Device Inoperable (1663); Protective Measures Problem (3015)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 09/12/2016 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).The actual device has been returned and is currently pending evaluation.Once reliability engineering evaluates the device, a supplemental medwatch report will be sent accordingly.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
|
Event Description
|
It was reported that during pre-surgery testing, it was discovered that the top trigger button was sticking and did not work on the small battery drive device.There were no delays to the surgical procedure as an identical spare device was available for use.There was no patient involvement reported.There were no reports of injuries, medical intervention or prolonged hospitalization.All available information has been disclosed.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
|
|
Manufacturer Narrative
|
Additional narrative:the actual device was returned for evaluation.Reliability engineering evaluated the device and observed that the top and bottom triggers were sticking.It was further observed that the device had worn trigger components.Therefore, the reported condition was confirmed.The assignable root cause was determined to be due to component wear from normal use over time.If information is obtained that was not available, a follow-up medwatch will be filed as appropriate.
|
|