• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CODMAN & SHURTLEFF, INC. CODMAN DISPOSABLE PERFORATOR; DRILLS, BURRS, TREPHINES & ACC.

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CODMAN & SHURTLEFF, INC. CODMAN DISPOSABLE PERFORATOR; DRILLS, BURRS, TREPHINES & ACC. Back to Search Results
Catalog Number 26-1221
Device Problem Unsealed Device Packaging (1444)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 08/30/2016
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Upon completion of the investigation a follow up report will be filed.
 
Event Description
We would like to report you nc products and ask you investigation and improvement for perforators.Recently we imported perforators(26-1221) and found that some pieces had primary package partially opened.We checked all imported pieces of this lot visually and found two pieces were affected.
 
Manufacturer Narrative
Upon completion of the investigation, it was noted that the observation found at incoming inspection in japan was confirmed.The two returned packages containing the perforators confirmed the complaint of "primary package partially opened".It appears that the plastic tray and lid were not properly sealed during the manufacturing/ packaging process.Based on the results of this investigation, no further action is required.Trends will be monitored for this and similar complaints.At the present time, this complaint is closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CODMAN DISPOSABLE PERFORATOR
Type of Device
DRILLS, BURRS, TREPHINES & ACC.
Manufacturer (Section D)
CODMAN & SHURTLEFF, INC.
new bedford industrial park
new bedford MA 02745
Manufacturer Contact
james kenney
325 paramount drive
raynham, MA 02767
5088282726
MDR Report Key5984635
MDR Text Key55880709
Report Number1226348-2016-10691
Device Sequence Number1
Product Code HBF
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K791101
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Medical Equipment Company Technician/Representative
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number26-1221
Device Lot NumberHJ003S
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/09/2016
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 09/01/2016
Initial Date FDA Received09/28/2016
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received10/21/2016
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-