The insulin pump passed the displacement test, self test, unexpected restart error test and basic occlusion test.All operating currents are within specification.Off no power alarm functioned properly.No motor error alarm noted during testing.The insulin pump was unable to prime during prime test due to faulty force sensor resistor.Analysis was unable to perform the occlusion test and excessive no delivery test due to prime and fill anomaly.The motor passed motor test.The insulin pump had a scratched display window, cracked battery tube threads, cracked reservoir tube, cracked reservoir tube lip and a missing end cap sticker.Note: this is a remediation mdr.Medtronic diabetes implemented revised mdr reportability criteria effective on july 1, 2014.Subsequently, medtronic diabetes conducted a one year retrospective review of complaints.This event was retrospectively identified to be reportable based on the revised mdr reportability criteria.
|