Failure analysis noted that the pump passed basic occlusion, prime/compromised force sensor system and displacement tests.The pump was stuck in a motor error alarm loop during bolus delivery and motor position encoder error alarm noted in the history file.The motor was tested outside of the device and passed.The motor may have had intermittent failure that was not detected during the testing.The displacement test functioned properly.There was no motion sensor test failure alarm.The pump had cracked reservoir tube lip, scratched display window and cracked case near the display window corners.Note: this is a remediation mdr.Medtronic diabetes implemented revised mdr reportability criteria effective on (b)(6) 2014.Subsequently, medtronic diabetes conducted a one year retrospective review of complaints.This event was retrospectively identified to be reportable based on the revised mdr reportability criteria.
|