MEDTRONIC PUERTO RICO OPERATIONS CO. SURESCAN; STIMULATOR, SPINAL-CORD, IMPLANTED (PAIN RELIEF) Back to Search Results
Model Number 97714
Device Problems Intermittent Continuity (1121); Migration or Expulsion of Device (1395); Unintended Collision (1429)
Patient Problem Therapeutic Effects, Unexpected (2099)
Event Date 07/01/2016
Event Type  malfunction  
Manufacturer Narrative
Concomitant medical products: product id: 977a260, serial# (b)(4), implanted: (b)(6) 2015, product type: lead.Product id: 977a260, serial# (b)(4), implanted: (b)(6) 2015, product type: lead.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event Description
The patient (via the manufacturer's representative) reported that the patient's stimulation was surging.The stimulation issue began after the patient had a serious fall in (b)(6) 2015.The manufacturer's representative (rep) reprogrammed the implantable neurostimulator (ins) and checked impedances.It was later reported that x-rays were taken to check the leads and the patient thought one of the leads had moved.No surgical intervention occurred.The indication for use was chronic low back pain and spinal pain.
 
Event Description
Additional information received stated that the reprogramming has eliminated the issue about the stimulating surging and the patient is doing well.
 
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SURESCAN
Type of Device
STIMULATOR, SPINAL-CORD, IMPLANTED (PAIN RELIEF)
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer Contact
diane wolf
7000 central avenue ne rcw215
minneapolis, MN 55432
7635263987
MDR Report Key5991809
MDR Text Key56102623
Report Number3004209178-2016-20169
Device Sequence Number1
Product Code GZB
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P840001
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,consum
Reporter Occupation Other
Type of Report Initial,Followup,Followup,Followup
Report Date 11/08/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/30/2016
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date10/28/2015
Device Model Number97714
Device Catalogue Number97714
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Date Manufacturer Received10/24/2016
Date Device Manufactured12/12/2014
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age62 YR
-
-