Catalog Number 102454407 |
Device Problem
Appropriate Term/Code Not Available (3191)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 09/28/2016 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.This complaint is still under investigation.Depuy will notify the fda of the results of this investigation once it has been completed.
|
|
Event Description
|
The surgeon was unable to set the poly insert in the tibial tray during surgery.
|
|
Manufacturer Narrative
|
If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.This complaint is still under investigation.Depuy will notify the fda of the results of this investigation once it has been completed.
|
|
Manufacturer Narrative
|
Examination of the submitted device noted scratches and gouges on the anterior portion of the distal surface, the anterior proximal edge, and the anterior medial side that are consistent with removal of the insert after unsuccessful attempts to mate the insert to the tray.Review of the device history records did not reveal any manufacturing deviations or anomalies.A worldwide complaint database search found no additional related reports against the complaint sample product code/lot code combination.Deformation observed on the device suggests that the root cause is misalignment of the insert relative to the tray during implantation.No evidence was found indicating product error was a contributing factor to the reported event and the need for corrective action is not indicated.Monitor complaints of assembly problems through sep-419.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
|
Search Alerts/Recalls
|