Catalog Number 102454407
Device Problem Appropriate Term/Code Not Available (3191)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 09/28/2016
Event Type  malfunction  
Manufacturer Narrative
(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.This complaint is still under investigation.Depuy will notify the fda of the results of this investigation once it has been completed.
 
Event Description
The surgeon was unable to set the poly insert in the tibial tray during surgery.
 
Manufacturer Narrative
If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.This complaint is still under investigation.Depuy will notify the fda of the results of this investigation once it has been completed.
 
Manufacturer Narrative
Examination of the submitted device noted scratches and gouges on the anterior portion of the distal surface, the anterior proximal edge, and the anterior medial side that are consistent with removal of the insert after unsuccessful attempts to mate the insert to the tray.Review of the device history records did not reveal any manufacturing deviations or anomalies.A worldwide complaint database search found no additional related reports against the complaint sample product code/lot code combination.Deformation observed on the device suggests that the root cause is misalignment of the insert relative to the tray during implantation.No evidence was found indicating product error was a contributing factor to the reported event and the need for corrective action is not indicated.Monitor complaints of assembly problems through sep-419.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SIGMA HP UNI INS SZ4 7MM RM/LL
Type of Device
KNEE TIBIAL BEARING/INSERT
Manufacturer (Section D)
DEPUY ORTHOPAEDICS, INC. 1818910
700 orthopaedic drive
warsaw IN 46582
Manufacturer (Section G)
DEPUY ORTHOPAEDICS, INC. 1818910
700 orthopaedic drive
warsaw IN 46582
Manufacturer Contact
chad gibson
700 orthopaedic drive
warsaw, IN 46581
5743725905
MDR Report Key5993404
MDR Text Key56181981
Report Number1818910-2016-28602
Device Sequence Number1
Product Code HRY
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K070267
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,health profession
Reporter Occupation Health Professional
Type of Report Initial,Followup,Followup
Report Date 09/28/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/03/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number102454407
Device Lot Number535315
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/06/2016
Is the Reporter a Health Professional? Yes
Date Manufacturer Received08/21/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/16/2014
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-