• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SYNCARDIA SYSTEMS, LLC SYNCARDIA CIRCULATORY SUPPORT SYSTEM CONSOLE; EXTERNAL PNEUMATIC DRIVER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SYNCARDIA SYSTEMS, LLC SYNCARDIA CIRCULATORY SUPPORT SYSTEM CONSOLE; EXTERNAL PNEUMATIC DRIVER Back to Search Results
Model Number CSS CONSOLE
Device Problems Break (1069); Image Display Error/Artifact (1304); Device Operates Differently Than Expected (2913); Device Operational Issue (2914)
Patient Problems No Patient Involvement (2645); No Known Impact Or Consequence To Patient (2692)
Event Date 09/21/2016
Event Type  malfunction  
Manufacturer Narrative
This alleged failure mode poses a low risk to a patient because the issue was observed when the css console was not supporting a patient.In addition, it would not prevent the css console from performing its life-sustaining functions.The computer is a monitoring device only and does not control css console functionality.The css console will be returned to syncardia for evaluation.The results of the investigation will be provided in a follow-up mdr.(b)(4).
 
Event Description
This css console was not supporting a patient.The customer reported that the monitoring laptop computer on the css console had four lines on the computer screen.The customer also reported that the css console had a broken rear left castor.
 
Manufacturer Narrative
The css console was returned to syncardia for evaluation.The reported issues, four lines on the monitor and a bad caster, were confirmed and reproduced during investigation testing.The monitor was found to have additional lines across the monitor during this investigation; however, the lcd screen still displayed all information visibly with normal contrast.The caster was found to have a damaged raceway but damage was insufficient to cause caster to improperly perform its functions and it continued performing as intended.This issue will continue to be monitored and trended as part of the customer experience process.Syncardia has completed its evaluation of this complaint and is closing this file.(b)(4).
 
Event Description
This css console was not supporting a patient.The customer reported that the monitoring laptop computer on the css console had four lines on the computer screen.The customer also reported that the css console had a broken rear left castor.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNCARDIA CIRCULATORY SUPPORT SYSTEM CONSOLE
Type of Device
EXTERNAL PNEUMATIC DRIVER
Manufacturer (Section D)
SYNCARDIA SYSTEMS, LLC
1992 e. silverlake rd.
tucson AZ 85713
Manufacturer Contact
michael garippa
1992 e. silverlake road
tucson, AZ 85713
5205451234
MDR Report Key5993557
MDR Text Key56807180
Report Number3003761017-2016-00327
Device Sequence Number1
Product Code LOZ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P030011
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 09/21/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberCSS CONSOLE
Device Catalogue Number400207
Device Lot Number1
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/05/2016
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 09/21/2016
Initial Date FDA Received10/03/2016
Supplement Dates Manufacturer Received01/22/2018
Supplement Dates FDA Received01/24/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/06/2016
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-