Model Number CSS CONSOLE |
Device Problems
Break (1069); Image Display Error/Artifact (1304); Device Operates Differently Than Expected (2913); Device Operational Issue (2914)
|
Patient Problems
No Patient Involvement (2645); No Known Impact Or Consequence To Patient (2692)
|
Event Date 09/21/2016 |
Event Type
malfunction
|
Manufacturer Narrative
|
This alleged failure mode poses a low risk to a patient because the issue was observed when the css console was not supporting a patient.In addition, it would not prevent the css console from performing its life-sustaining functions.The computer is a monitoring device only and does not control css console functionality.The css console will be returned to syncardia for evaluation.The results of the investigation will be provided in a follow-up mdr.(b)(4).
|
|
Event Description
|
This css console was not supporting a patient.The customer reported that the monitoring laptop computer on the css console had four lines on the computer screen.The customer also reported that the css console had a broken rear left castor.
|
|
Manufacturer Narrative
|
The css console was returned to syncardia for evaluation.The reported issues, four lines on the monitor and a bad caster, were confirmed and reproduced during investigation testing.The monitor was found to have additional lines across the monitor during this investigation; however, the lcd screen still displayed all information visibly with normal contrast.The caster was found to have a damaged raceway but damage was insufficient to cause caster to improperly perform its functions and it continued performing as intended.This issue will continue to be monitored and trended as part of the customer experience process.Syncardia has completed its evaluation of this complaint and is closing this file.(b)(4).
|
|
Event Description
|
This css console was not supporting a patient.The customer reported that the monitoring laptop computer on the css console had four lines on the computer screen.The customer also reported that the css console had a broken rear left castor.
|
|