U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC PUERTO RICO OPERATIONS CO. PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP; INSULIN INFUSION PUMP / SENSOR AUGMENTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC PUERTO RICO OPERATIONS CO. PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP; INSULIN INFUSION PUMP / SENSOR AUGMENTED Back to Search Results
Model Number MMT-523NAS
Device Problems Device Displays Incorrect Message (2591); Electromagnetic Compatibility Problem (2927)
Patient Problem Hyperglycemia (1905)
Event Date 02/06/2014
Event Type  malfunction  
Manufacturer Narrative
Failure analysis noted that the pump passed rewind, basic occlusion and displacement tests.The pump was stuck in a motor error alarm loop during bolus delivery and motor position encoder error alarm noted in the history file.The motor was tested outside of the device and passed.The motor may have had intermittent failure that was not detected during the testing.The pump had scratched lcd window and cracked reservoir tube lip.Note: this is a remediation mdr.Medtronic diabetes implemented revised mdr reportability criteria effective on july 1, 2014.Subsequently, medtronic diabetes conducted a one year retrospective review of complaints.This event was retrospectively identified to be reportable based on the revised mdr reportability criteria.(b)(4).
 
Event Description
The customer reported via phone call that the insulin pump had motor error alarm and motor position encoder error alarm.The blood glucose at the time of the incident was 169 mg/dl.The customer was changing the infusion set when the motor error alarm was received.The drive support disk was normal.The customer had an x-ray with the pump on.They were able to rewind the pump.Troubleshooting was performed.The device was returned for analysis.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP
Type of Device
INSULIN INFUSION PUMP / SENSOR AUGMENTED
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50
road 31 km 24.4
juncos PR 00777 3869
Manufacturer (Section G)
MEDTRONIC MINIMED
18000 devonshire street
northridge CA 91325 1219
Manufacturer Contact
gerwin de graaff
18000 devonshire street
northridge, CA 91325-1219
8185764805
MDR Report Key6003617
MDR Text Key57186921
Report Number2032227-2016-26441
Device Sequence Number1
Product Code OYC
Combination Product (y/n)N
PMA/PMN Number
P980022
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Reporter Occupation Patient
Type of Report Initial
Report Date 02/06/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberMMT-523NAS
Device Catalogue NumberMMT-523NAS
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/13/2014
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 02/06/2014
Initial Date FDA Received10/06/2016
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Age53 YR
Patient Weight79
-
-