Failure analysis noted that the pump passed rewind, basic occlusion and displacement tests.The pump was stuck in a motor error alarm loop during bolus delivery and motor position encoder error alarm noted in the history file.The motor was tested outside of the device and passed.The motor may have had intermittent failure that was not detected during the testing.The pump had scratched lcd window and cracked reservoir tube lip.Note: this is a remediation mdr.Medtronic diabetes implemented revised mdr reportability criteria effective on july 1, 2014.Subsequently, medtronic diabetes conducted a one year retrospective review of complaints.This event was retrospectively identified to be reportable based on the revised mdr reportability criteria.(b)(4).
|