Model Number D314TRG
Device Problems False Alarm (1013); Electromagnetic Interference (1194); Over-Sensing (1438)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 07/25/2016
Event Type  Injury  
Manufacturer Narrative
Concomitant product(s): a 4543 boston scientific lead, implanted : (b)(6) 2009.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that a lead integrity alert (lia) triggered on the right ventricular (rv) lead for oversensing with sensing integrity counter (sic) and noise.However, the electrograms were most consistent with electromagnetic interference (emi) exposure of the implantable cardioverter defibrillator (icd) device and not a lead integrity issue.Reprogramming was initially performed, but the device was later explanted and replaced.No patient complications have been reported as a result of this event.
 
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROTECTA XT CRT-D
Type of Device
DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CARDIAC RESYNCHRONIZATIO
Manufacturer (Section D)
MEDTRONIC MED REL MEDTRONIC PUERTO RICO
ceiba norte industrial park, r
juncos PR 00777
Manufacturer (Section G)
MEDTRONIC MED REL MEDTRONIC PUERTO RICO
ceiba norte industrial park, r
juncos PR 00777
Manufacturer Contact
anne schilling
8200 coral sea st ne
mounds view, MN 55112
7635052036
MDR Report Key6005411
MDR Text Key56763199
Report Number3004209178-2016-20665
Device Sequence Number1
Product Code NIK
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P010031
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 07/28/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date08/28/2013
Device Model NumberD314TRG
Device Catalogue NumberD314TRG
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 07/28/2016
Initial Date FDA Received10/06/2016
Supplement Dates Manufacturer Received07/28/2016
Supplement Dates FDA Received09/25/2017
Date Device Manufactured03/12/2012
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
694765 LEAD, 5568-53 LEAD
Patient Outcome(s) Hospitalization; Required Intervention;
Patient Age85 YR
-
-