VENTLAB LLC. AIRFLOW; RESUS BAG Back to Search Results
Model Number AF5142MBP
Device Problems Device Issue (2379); Activation, Positioning or Separation Problem (2906)
Patient Problems Not Applicable (3189); Missing Value Reason (3192)
Event Date 09/09/2016
Event Type  malfunction  
Manufacturer Narrative
An extensive and thorough investigation was performed by the engineering team that concluded the root cause for this problem is the fact that the patient port and the mask are made out of similar base materials resulting in an undesirably high coefficient of friction between the two parts.Steps have been taken to implement a material combination with a lower coefficient of friction that significantly reduces the effort required to remove the mask from the patient valve.
 
Event Description
The customer alleges "mask was sticking to the bag and the patient valve broke off".No other details were provided and no patient injury/harm reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AIRFLOW
Type of Device
RESUS BAG
Manufacturer (Section D)
VENTLAB LLC.
2710 northridgedr. suite a
grand rapids MI 49544
Manufacturer (Section G)
VENTLAB LLC.
2710 northridgedr. suite a
grand rapids MI 49544
Manufacturer Contact
carrie fortuna
2710 northridge dr. suite a
grand rapids, MI 49544
6162598350
MDR Report Key6010529
MDR Text Key56794280
Report Number2246980-2016-00053
Device Sequence Number1
Product Code BTM
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K012842
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor
Reporter Occupation Other
Type of Report Initial
Report Date 10/07/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Model NumberAF5142MBP
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 09/09/2016
Initial Date FDA Received10/07/2016
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Type of Device Usage N
Patient Sequence Number1
-
-