Catalog Number 06437281190 |
Device Problem
High Test Results (2457)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 09/05/2016 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).This event occurred in (b)(6).
|
|
Event Description
|
The customer received questionable elecsys ft4 ii assay and elecsys tsh assay results for two samples from on patient from two cobas e602 analyzers and suspected the possible presence of heterophilic antibodies.Refer to the attachment to the medwatch for all patient data.This medwatch is for the ft4 assay.Refer to the medwatch with (b)(6) for the tsh assay.The results were reported to the clinicians.The patient was not adversely affected.As no sample material from the patient could be provided for further investigation, a specific root cause could not be determined.From the provided information, a general reagent issue was not detected.
|
|
Manufacturer Narrative
|
A new sample was collected from the patient on (b)(6) 2016 and generated questionable results.
|
|
Manufacturer Narrative
|
Two samples from the patient were submitted for investigation.A specific root cause could not be determined.The testing confirmed the customer's results as the ft4 results generated were above the reference range.No interfering factors were detected in the samples.For diagnostic purposes, the results should always be assessed win conjunction with the patient's medical history, clinical examination, and other findings.
|
|
Search Alerts/Recalls
|