• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC CRYOCATH LP FLEXCATH ADVANCE STEERABLE SHEATH; CATHETER, STEERABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC CRYOCATH LP FLEXCATH ADVANCE STEERABLE SHEATH; CATHETER, STEERABLE Back to Search Results
Model Number 4FC12
Device Problem Air Leak (1008)
Patient Problems Air Embolism (1697); ST Segment Elevation (2059)
Event Date 09/09/2016
Event Type  Injury  
Manufacturer Narrative
A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that during a cryo ablation procedure air introduction into the patient's left atrium was alleged.Procedure outcome was unknown at this time.No further patient complications have been reported as a result of this event.
 
Manufacturer Narrative
The sheath was returned and analyzed.Visual inspection of the sheath showed that the shaft was intact with no apparent issues.The reported air ingress could not be reproduced.Multiple aspirations and injections were performed without air bubbles or leaks.Also, the hemostatic valve and the valve assembly were leak tight.In conclusion, the reported air ingress issue was not confirmed through testing.A clinical issue of st elevation was encountered during the procedure.The sheath passed the returned product inspection as per specification.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was further reported that air was not visible and "just the st elevation was observed on the ecg that took some time to resolve".No further patient complications have been reported as a result of this event.
 
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FLEXCATH ADVANCE STEERABLE SHEATH
Type of Device
CATHETER, STEERABLE
Manufacturer (Section D)
MEDTRONIC CRYOCATH LP
16771 chemin ste-marie
kirkland H9H 5 H3
CA  H9H 5H3
Manufacturer (Section G)
MEDTRONIC CRYOCATH LP
16771 chemin ste-marie
kirkland H9H 5 H3
CA   H9H 5H3
Manufacturer Contact
anne schilling
8200 coral sea st ne
mounds view, MN 55112
7635052036
MDR Report Key6010900
MDR Text Key56801962
Report Number3002648230-2016-00427
Device Sequence Number1
Product Code DRA
Combination Product (y/n)N
Reporter Country CodeSF
PMA/PMN Number
K123591
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial,Followup,Followup,Followup
Report Date 09/09/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date04/28/2018
Device Model Number4FC12
Device Catalogue Number4FC12
Device Lot Number24900
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/23/2016
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 09/09/2016
Initial Date FDA Received10/07/2016
Supplement Dates Manufacturer ReceivedNot provided
Not provided
12/19/2016
Supplement Dates FDA Received11/08/2016
12/19/2016
09/14/2017
Was Device Evaluated by Manufacturer? No
Date Device Manufactured04/29/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
2AF283 CATHETER
Patient Outcome(s) Other;
-
-