The insulin pump passed displacement test, rewind, basic occlusion and prime/ compromised force sensor tests.The pump was received stuck in motor error alarm loop during bolus/basal delivery.The motor was tested outside of the device and passed.The motor may have had an intermittent failure that was not detected during our testing.Analysis was unable to confirm motor position encoder error alarm due to history file was overwritten.The pump had a cracked display window corner, scratched lcd window, cracked battery tube threads and cracked reservoir tube lip.Note: this is a remediation mdr.Medtronic diabetes implemented revised mdr reportability criteria effective on july 1, 2014.Subsequently, medtronic diabetes conducted a one year retrospective review of complaints.This event was retrospectively identified to be reportable based on the revised mdr reportability criteria.(b)(4).
|
The customer reported via phone call that the pump had motor error alarm and motor position encoder error alarm.The blood glucose at the time of the incident was 132 mg/dl.The customer states the pump was exposed to a high magnetic field, during a cat scan, two prior to the call.The drive support cap appears intact.The customer stated there was a motor position encoder error alarm noted in the history.The customer states they were not able to rewind the pump.Troubleshooting was not able to resolve the issue.The customer was advised to disconnect from the insulin pump and revert to back-up plan.The customer was advised that the insulin pump will need to be replaced.The device will be returned for analysis.
|