Model Number AG-920RA
Device Problems Device Inoperable (1663); Device Displays Incorrect Message (2591)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 09/07/2016
Event Type  malfunction  
Manufacturer Narrative
The customer reported that the multi-gas unit stopped working, sensor calibration error carbon-dioxide.Nihon kohden continues to investigate the reported event.Nihon kohden will submit a supplemental report in accordance with 21 cfr section 803.56 when additional information becomes available.
 
Event Description
The customer reported that the multi-gas unit stopped working, sensor calibration error carbon-dioxide.
 
Manufacturer Narrative
Manufacturer narrative: the customer reported that the multi-gas unit stopped working, sensor calibration error carbon-dioxide.The unit was cleaned and evaluated.The reported problem of intermittent cal error was duplicated.The gas sensing unit was replaced to resolve the issue.The device has been repaired and returned to the customer.Nihon kohden will submit a supplemental report in accordance with 21 cfr part 803.56 if additional information becomes available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AG-920RA
Type of Device
MULTI-GAS UNIT
Manufacturer (Section D)
NIHON KOHDEN CORPORATION
1-31-4 nishiochia, shinjuku-ku
attn: shama mooman
tokyo, japan 161-8 560
JA  161-8560
Manufacturer (Section G)
NIHON KOHDEN TOMIOKA CORPORATION
1-1 tajino
attn: shama mooman
gunma 370-2 314
JA   370-2314
Manufacturer Contact
shama mooman
safety mgmt dept, quality mgmt
seibu bldg 2, 4th floor 1-11-2
kusunokidai tokorozawa, saitama 359-8-580
JA   359-8580
2687488
MDR Report Key6012556
MDR Text Key57317075
Report Number8030229-2016-00510
Device Sequence Number1
Product Code CCK
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K020046
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date 10/07/2016,09/07/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/07/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberAG-920RA
Device Catalogue NumberAG-920RA
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/17/2016
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? Yes
Date Report Sent to FDA10/07/2016
Distributor Facility Aware Date09/07/2016
Event Location Hospital
Date Report to Manufacturer10/07/2016
Date Manufacturer Received10/07/2016
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-