MAUDE Adverse Event Report: SORIN GROUP DEUTSCHLAND SORIN CENTRIFUGAL PUMP 5 (CP5); CONTROL, PUMP SPEED, CARDIOPULMONARY BYPASS,

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SORIN GROUP DEUTSCHLAND SORIN CENTRIFUGAL PUMP 5 (CP5); CONTROL, PUMP SPEED, CARDIOPULMONARY BYPASS, Back to Search Results
Model Number 60-01-04
Device Problems Device Operational Issue (2914); No Flow (2991); No Pressure (2994)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 09/07/2016
Event Type  No Answer Provided  
Manufacturer Narrative
On (b)(4) 2016, sorin group received a user medwatch report (mw5064715) regarding this event.However, the user report was filed against the disposable and not the equipment.Please reference medwatch report number 9680841-2016-00503 for information on the disposable.Sorin group (b)(4) manufactures the sorin centrifugal pump 5 (cp5).The incident occurred in (b)(6).This medwatch report is filed on behalf of sorin group (b)(4).Sorin group (b)(4) received a report that the cardioplegia delivery of the sorin centrifugal pump 5 (cp5) lost flow and almost all pressure for about 15 minutes.The entire circuit and heart-lung machine were switched out to continue the procedure.The case was completed without further incident.There was no report of patient injury.A review of the dhr did not identify any deviations or non-conformities relevant to the reported issue.The investigation is on-going.A follow-up report will be sent when the investigation is complete.
 
Event Description
Sorin group (b)(4) received a report that the cardioplegia delivery of the sorin centrifugal pump 5 (cp5) lost flow and almost all pressure for about 15 minutes.The entire circuit and heart-lung machine were switched out to continue the procedure.The case was completed without further incident.There was no report of patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SORIN CENTRIFUGAL PUMP 5 (CP5)
Type of Device
CONTROL, PUMP SPEED, CARDIOPULMONARY BYPASS,
Manufacturer (Section D)
SORIN GROUP DEUTSCHLAND
lindberghstr. 25
munich, 80939
GM  80939
Manufacturer (Section G)
SORIN GROUP DEUTSCHLAND
lindberghstr. 25
munich, 80939
GM   80939
Manufacturer Contact
carrie wood
14401 w. 65th way
arvada, CO 80004
3034676461
MDR Report Key6012759
MDR Text Key56973644
Report Number9611109-2016-00707
Device Sequence Number1
Product Code DWA
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K112225
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Health Professional
Remedial Action Other
Type of Report Initial
Report Date 09/07/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/07/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number60-01-04
Is the Reporter a Health Professional? Yes
Date Manufacturer Received09/07/2016
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured08/06/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-