MAUDE Adverse Event Report: GUIDANT CRM CLONMEL IRELAND VITALITY 2; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

GUIDANT CRM CLONMEL IRELAND VITALITY 2; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR Back to Search Results
Model Number T175
Device Problems Pacing Problem (1439); Device Sensing Problem (2917)
Patient Problem Dizziness (2194)
Event Date 03/05/2012
Event Type  Injury  
Manufacturer Narrative
Upon receipt at our post market quality assurance laboratory, a thorough evaluation of the device was performed.A visual inspection of the device header and case noted no anomalies.Pin gauge testing, designed to verify proper port dimensions, was completed.Each port measured as expected.The device was then exposed to simulated heart load conditions, and the defibrillation, pacing, and sensing functions were tested.Impedance testing was completed and all measurements were within normal limits.The device operated appropriately with no interruptions in therapy output at the returned programmed settings.A series of electrical tests was also performed, and again, normal device function was observed.
 
Event Description
Boston scientific received information that the patient with this implantable cardioverter defibrillator (icd) presented to the emergency room for dizziness and lightheadedness.The patient's ventricular rate was inconsistent and the field representative contacted a technical services (ts) consultant to discuss programming the rate smoothing function.It was noted that pacing inhibition was causing the symptoms.The duration of the pacing inhibition is unknown at this time.Additional information indicated that there was no pacing inhibition greater than 2 seconds.The lower rate limit was programmed from 40 to 50 to resolve the issue.This device was subsequently explanted.No additional adverse patient effects were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VITALITY 2
Type of Device
IMPLANTABLE CARDIOVERTER DEFIBRILLATOR
Manufacturer (Section D)
GUIDANT CRM CLONMEL IRELAND
guidant ireland
clonmel, tipperary ireland
Manufacturer (Section G)
GUIDANT CRM CLONMEL IRELAND
guidant ireland
clonmel, tipperary ireland
Manufacturer Contact
sonali vasekar
4100 hamline ave. n
st. paul, MN 
6515824786
MDR Report Key6021230
MDR Text Key57153608
Report Number2124215-2016-13235
Device Sequence Number1
Product Code KRG
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Medical Equipment Company Technician/Representative
Type of Report Initial
Report Date 07/28/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/12/2016
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date01/16/2009
Device Model NumberT175
Other Device ID NumberVITALITY 2 VR
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/30/2016
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? Yes
Date Manufacturer Received07/28/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/17/2008
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
0158; 1850; T175
Patient Outcome(s) Hospitalization; Life Threatening; Required Intervention;
Patient Age79 YR
-
-