SYNTHES BRANDYWINE DEPTH GAUGE FOR 2.0MM AND 2.4MM SCREWS; GAUGE, DEPTH Back to Search Results
Catalog Number 319.006
Device Problem Break (1069)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
It is unknown if there was patient involvement.Device is an instrument and is not implanted/explanted.Complainant part is expected to be returned for manufacturer review/investigation, but has yet to be received.No service history review can be performed as part number 319.006 with lot number(s) 5252431 is a lot/batch controlled item.The manufacture date of this item is (b)(6) 2006.The source of the manufacture date is the release to warehouse date.The service history review is unconfirmed.The investigation could not be completed; no conclusion could be drawn, as no product was received.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
It was reported that a depth gauge has a broken needle.It is unknown how and where this occurred.This is report 1 of 1 for (b)(4).
 
Manufacturer Narrative
Part 319.006, lot 5252431: release to warehouse date: june 08, 2006.Manufactured by synthes (b)(4).Review of the device history records (dhr) showed there are potential issues during the manufacture of the product that would contribute to this complaint condition.A non-conformance report was written for various lots including 5252431 an undersized major diameter of the thread feature on the needle component.As the complaint description notes that the needle failed, relevance to the complaint condition cannot be determined until the product is returned for investigation; however, the bending strength of an external thread element would be a function of the size of the minor diameter, not the major diameter since the minor diameter represents the worst case cross section in terms of bending strength.All subcomponent dhrs were reviewed and found to be without issue during the manufacturing process that would contribute to this complaint condition.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DEPTH GAUGE FOR 2.0MM AND 2.4MM SCREWS
Type of Device
GAUGE, DEPTH
Manufacturer (Section D)
SYNTHES BRANDYWINE
1303 goshen parkway
west chester PA 19380
Manufacturer (Section G)
SYNTHES BRANDYWINE
1303 goshen parkway
west chester PA 19380
Manufacturer Contact
mark vornheder
1302 wrights lane east
west chester, PA 19380
6107195000
MDR Report Key6022484
MDR Text Key57272829
Report Number2530088-2016-10287
Device Sequence Number1
Product Code HTJ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Health Professional
Type of Report Initial,Followup
Report Date 09/22/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/12/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number319.006
Device Lot Number5252431
Other Device ID Number(01)10886982189943(10)5252431
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received10/25/2016
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured06/08/2006
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-