Model Number PB1018 |
Device Problems
Incomplete Coaptation (2507); Device Operates Differently Than Expected (2913)
|
Patient Problem
Tricuspid Regurgitation (2112)
|
Event Date 09/07/2016 |
Event Type
Injury
|
Manufacturer Narrative
|
Without the return of the product, no definitive conclusion can be made regarding the clinical observation.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
Medtronic received information that during the implant of this transcatheter pulmonary bioprosthetic valve (tpbv) in the tricuspid position the valve showed moderate central regurgitation and a second valve was implanted valve-in-valve which resolved the regurgitation.No adverse patient effects were reported.The physician reported: "one of the leaflets of the melody valve was riding up to a different plane than the other two leaflets, creating a coaptation defect and a very eccentric, almost horizontal jet from the plane of the tricuspid valve".
|
|
Manufacturer Narrative
|
The device history record was reviewed and showed that this product met all manufacturing specifications for product released for distribution.No issues were identified that would have impacted this event.The melody valve is not designed for the tricuspid position.The device was not returned for analysis.No root cause can be determined at this time.
|
|
Manufacturer Narrative
|
If information is provided in the future, a supplemental report will be issued.
|
|