Model Number PB1018
Device Problems Incomplete Coaptation (2507); Device Operates Differently Than Expected (2913)
Patient Problem Tricuspid Regurgitation (2112)
Event Date 09/07/2016
Event Type  Injury  
Manufacturer Narrative
Without the return of the product, no definitive conclusion can be made regarding the clinical observation.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event Description
Medtronic received information that during the implant of this transcatheter pulmonary bioprosthetic valve (tpbv) in the tricuspid position the valve showed moderate central regurgitation and a second valve was implanted valve-in-valve which resolved the regurgitation.No adverse patient effects were reported.The physician reported: "one of the leaflets of the melody valve was riding up to a different plane than the other two leaflets, creating a coaptation defect and a very eccentric, almost horizontal jet from the plane of the tricuspid valve".
 
Manufacturer Narrative
The device history record was reviewed and showed that this product met all manufacturing specifications for product released for distribution.No issues were identified that would have impacted this event.The melody valve is not designed for the tricuspid position.The device was not returned for analysis.No root cause can be determined at this time.
 
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MELODY TRANSCATHER PULMONARY
Type of Device
PULMONARY VALVE PROSTHESIS PERCUTANEOUSLY DELIVERED
Manufacturer (Section D)
MEDTRONIC HEART VALVE DIVISION
1851 e deere ave
santa ana CA 92705
Manufacturer (Section G)
MEDTRONIC HEART VALVE DIVISION
1851 e deere ave
santa ana CA 92705
Manufacturer Contact
paula bixby
8200 coral sea street ne
mounds view, MN 55112
7635055378
MDR Report Key6026292
MDR Text Key57305423
Report Number2025587-2016-01614
Device Sequence Number1
Product Code NPV
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
H080002
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial,Followup,Followup,Followup
Report Date 12/29/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/13/2016
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date07/12/2018
Device Model NumberPB1018
Device Catalogue NumberPB1018
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received12/29/2016
Was Device Evaluated by Manufacturer? No
Date Device Manufactured07/28/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age58 YR
-
-