Current information is insufficient to permit a conclusion as to the cause of the event.Review of device history records show the lot released with no recorded anomaly or deviation.Examination of manufacturing history found no evidence of product non-conformance and device likely left the manufacturer conforming to print.Visual inspection of the drivers confirms both are deformed at the distal tip; however, the driver tip twisting is an intentional part of the driver's design.The twisting prevents the screw head from fracturing or stripping.A conclusive root cause of the event could not be determined with the available information; however, the most likely root cause was application of torque higher than the driver is designed to withstand.There are warnings in the package insert that state that this type of event can occur: under care and handling of instruments, it states, "surgical instruments and instrument cases are susceptible to damage for a variety of reasons including prolonged use, misuse, rough or improper handling.Care must be taken to avoid compromising their exacting performance." under precautions, it states, "biomet recommends that all instruments be regularly inspected for wear and disfigurement prior to use.".
|