Model Number 701028718
Device Problem Loss of Power (1475)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 09/21/2016
Event Type  malfunction  
Manufacturer Narrative
Maquet medical systems, usa submits this report on behalf of the legal manufacturer of the device maquet cardiopulmonary (b)(4).A maquet field service technician was on site and investigated the unit.The technician.Powered the unit on and after 5 minutes the unit shut off.The technician found the 12a breaker tripped.Re-set.Tested with customers hoses at multiple temperature settings running both pumps 1.5 hours.Unable to duplicate complaint.The technician connected the laptop with w/pc load.Verified current calibration within factory specifications.Performed functional and safety tests.A supplemental medwatch will be submitted if additional information becomes available.
 
Event Description
"according to the customer: the customer stated that the hcu30 shut down while on patient.Additional information: there were no negative consequences for the patient reported (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MAQUET HCU30 DEVICE
Type of Device
CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
MAQUET CARDIOPULMONARY AG
rastatt
GM 
Manufacturer (Section G)
BERND RAKOW
maquet cardiopulmonary ag
kehler strasse 31
76437 rastatt
GM  
Manufacturer Contact
maquet cardiopulmonary ag
kehler strasse 31
76437 rastatt 
4972229321
MDR Report Key6026561
MDR Text Key57695353
Report Number8010762-2016-00626
Device Sequence Number1
Product Code DWC
Combination Product (y/n)N
PMA/PMN Number
K031544
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer,health professional
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 09/21/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/13/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number701028718
Device Catalogue Number939001
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received09/21/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/01/2006
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-