Maquet medical systems, usa submits this report on behalf of the legal manufacturer of the device maquet cardiopulmonary (b)(4).A maquet field service technician was on site and investigated the unit.The technician.Powered the unit on and after 5 minutes the unit shut off.The technician found the 12a breaker tripped.Re-set.Tested with customers hoses at multiple temperature settings running both pumps 1.5 hours.Unable to duplicate complaint.The technician connected the laptop with w/pc load.Verified current calibration within factory specifications.Performed functional and safety tests.A supplemental medwatch will be submitted if additional information becomes available.
|