Model Number 195240
Device Problem Failure To Adhere Or Bond (1031)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 09/19/2016
Event Type  malfunction  
Manufacturer Narrative
(b)(4).The customer is using a medtronic reservoir.They have been using the level sensor for decades.
 
Event Description
It was reported that during set-up of the device for a cardiopulmonary bypass procedure, the level sensor pad is not sticking and they had a hard time getting it to stick on the reservoir.The product was switched out with another level sensor.The surgical procedure was completed successfully.There was no delay, no blood loss, nor adverse consequences to the patient.
 
Manufacturer Narrative
During laboratory analysis, the product surveillance technician (pst) the two returned pads stayed firmly in place on an oxygenator with level sensors mounted to the pads.No ¿not attached¿ alarms occurred while the level sensors were mounted to the oxygenator and connected to a level detect module.If additional information becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TERUMO ADVANCED PERFUSION SYSTEM 1
Type of Device
CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS-ADVANCED PERFUSION SYSTEM 1
Manufacturer (Section D)
TERUMO CARDIOVASCULAR SYSTEMS CORP.
6200 jackson road
ann arbor MI 48103
Manufacturer Contact
jan winder
6200 jackson road
ann arbor, MI 48103
7346634145
MDR Report Key6028871
MDR Text Key57622078
Report Number1828100-2016-00677
Device Sequence Number1
Product Code DTQ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K022947
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 11/28/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date02/22/2018
Device Model Number195240
Device Catalogue Number195240
Device Lot Number800015
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/03/2016
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 09/20/2016
Initial Date FDA Received10/13/2016
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received11/28/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/06/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
MEDTRONIC RESERVOIR
-
-