• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DENTSPLY TULSA DENTAL SPECIALTIES X-TIP INTRAOSSEOUS ANESTHESIA DELIVERY SYSTEM; NEEDLE, DENTAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DENTSPLY TULSA DENTAL SPECIALTIES X-TIP INTRAOSSEOUS ANESTHESIA DELIVERY SYSTEM; NEEDLE, DENTAL Back to Search Results
Catalog Number XTIP50
Device Problem Break (1069)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Type  malfunction  
Manufacturer Narrative
There has been a previous report received where this malfunction resulted in a serious injury.Therefore, it must be presumed that recurrence of this malfunction could possibly cause or contribute to a serious injury or require medical or surgical intervention to preclude such.As such, this event is reportable per 21 cfr part 803.This report is for the third device.The device is available for evaluation, though results are not available as of this report.Evaluation results will be submitted as they become available.
 
Event Description
It was reported that three x-tips broke during use; no injury resulted.
 
Manufacturer Narrative
Customer returned 50 unopened drill and guide sleeves for lot number 115742 and 6 up opened drill and guide sleeves from previous lot number that had expired on 05 15.No bent, broken, or opened product was returned.Unknown which lot customer was referring to in the complaint.No testing performed on expired product.Using the ansi plan with and aql of 1.5 using normal sample plan randomly selected and tested 8 pieces from the lot number with non expired product.Using microscope visually checked drills.No manufacturing defects or markings noted on drills.Using test method tm-0009 for the go ring gauge.All 8 tested passed the through the go ring gauge.Tested drills using a push button contra angle and drills fit into the contra angle with no hesitation.No fault found with returned drills.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
X-TIP INTRAOSSEOUS ANESTHESIA DELIVERY SYSTEM
Type of Device
NEEDLE, DENTAL
Manufacturer (Section D)
DENTSPLY TULSA DENTAL SPECIALTIES
608 rolling hills drive
johnson city TN 37604
Manufacturer (Section G)
DENTSPLY TULSA DENTAL SPECIALTIES
608 rolling hills drive
johnson city TN 37604
Manufacturer Contact
helen lewis
221 w. philadelphia st.
suite 60w
york, PA 17401
7178494229
MDR Report Key6031575
MDR Text Key57717641
Report Number2320721-2016-00056
Device Sequence Number1
Product Code DZM
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K990511
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Dentist
Type of Report Initial,Followup
Report Date 01/18/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/14/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberXTIP50
Device Lot Number0000115742
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/06/2016
Is the Reporter a Health Professional? Yes
Date Manufacturer Received12/19/2016
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-