(b)(4).We received one used, empty easypump ii lt 125-25-s without packaging.The provided sample was subjected to a visual inspection.Damages or manufacturing faults were not detected.As-received condition the clamp clip was closed.The patient connector was not closed, the original wing cap was not handed over from the customer.Further on, we detected crystallized drug residues and solution at the filling port (lli-cone) of the sample.Additionally, the pump was filled with nacl 0.9 % up to the nominal volume (125 ml) and a functional test respectively.A leak test was carried out.After opening the clamp clip and starting the pump the pump did work immediately (solution was running).Leakages were not detected at the sample.Flow rate test: nominal: 5 ml/h.Actual: 7.1 ml in 1 h; 12.3 ml in 2 hrs; 65.9 ml in 16 hrs.A fast flow (as described by the customer) could not be determined at the received sample.Device history records (dhr): reviewed device history records, there is no abnormalities and no such defect detected at final control inspection.All available information has been forwarded to the actual manufacturer.If additional pertinent information becomes available, a follow up report will be filed.
|