• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC.(CRM-SUNNYVALE) QUARTET; PERMANENT PACEMAKER ELECTRODE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC.(CRM-SUNNYVALE) QUARTET; PERMANENT PACEMAKER ELECTRODE Back to Search Results
Model Number 1458Q/86
Device Problem High Capture Threshold (3266)
Patient Problem Syncope (1610)
Event Date 09/09/2016
Event Type  Injury  
Manufacturer Narrative
(b)(4).
 
Event Description
It was reported that patient presented to hospital with syncope.Upon device interrogation, there were no episodes noted to explain syncope, but loss of capture was observed on the lv lead.Lead repositioning is planned in the near future but not yet scheduled.The device was reprogrammed.
 
Manufacturer Narrative
(b)(4).
 
Event Description
New information received indicated that the lead was explanted and replaced.The patient presented to the cath lab for lead repositioning and x-ray was performed showing the lead had dislodged into the cs with a u-shape in the vein.Physician felt the lead was not reusable and was cut to advance a wire through to preserve position in cs, the lead was removed.The patient tolerated the procedure well.
 
Manufacturer Narrative
The damage found was sustained during the surgical procedure.The lead was otherwise normal.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUARTET
Type of Device
PERMANENT PACEMAKER ELECTRODE
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SUNNYVALE)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SUNNYVALE)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
jaime chavez
645 almanor avenue
sunnyvale, CA 94085
8184934022
MDR Report Key6042710
MDR Text Key57918936
Report Number2938836-2016-13037
Device Sequence Number1
Product Code OJX
Combination Product (y/n)N
Reporter Country CodeCA
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Nurse
Type of Report Initial,Followup,Followup
Report Date 09/09/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/19/2016
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date12/31/2014
Device Model Number1458Q/86
Device Lot Number3580672
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/08/2016
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/18/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/15/2011
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age61 YR
-
-