MEDTRONIC PUERTO RICO OPERATIONS CO. ENTERRA; INTESTINAL STIMULATOR Back to Search Results
Model Number 37800
Device Problem Break (1069)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 09/21/2016
Event Type  Injury  
Manufacturer Narrative
Concomitant products: product id: 4351, product type: lead.
 
Event Description
The health care provider (hcp) via a manufacturer representative reported that the lead was damaged during the implant procedure.The prolene was damaged due to the stapler that the hcp used to staple down the leads while they sutured the leads to the stomach wall.The manufacturer representative believed this was likely due to user error since this was the hcp's first implant.There was an insulation breach and the insulation was damaged.The hcp implanted another new lead.The lead would be coming back in for analysis.The manufacturer representative didn't really think anything was wrong with the lead, just the possible damage done at implant.The damaged replacement lead issue had been resolved.No medical symptoms were reported.The indications for use for this patient were gastric stimulation and gastrointestinal/pelvic floor.
 
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENTERRA
Type of Device
INTESTINAL STIMULATOR
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer Contact
diane wolf
7000 central avenue ne rcw215
minneapolis, MN 55432
7635263987
MDR Report Key6044742
MDR Text Key57974872
Report Number3004209178-2016-22201
Device Sequence Number1
Product Code LNQ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
H990014
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 10/20/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/20/2016
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date01/14/2018
Device Model Number37800
Device Catalogue Number37800
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received09/21/2016
Date Device Manufactured07/26/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age52 YR
-
-