Catalog Number 004551004
Device Problems Break (1069); Device Damaged Prior to Use (2284)
Patient Problem No Patient Involvement (2645)
Event Date 10/07/2016
Event Type  malfunction  
Manufacturer Narrative
(b)(4).The investigation is in process at the time of this report.
 
Event Description
Customer alleges laryngoscope blade received broken in the package at receipt.There was no reported patient involvement.
 
Manufacturer Narrative
(b)(4).The device history record was reviewed and no issue that could have contributed to the reported failure was noted.The device was manufactured according to release specification.The sample was returned for evaluation.A visual exam was performed and it was observed that the blade was broken.It is unknown where the damage occurred as all products are inspected at the manufacturing facility prior to shipment; therefore, a defect of this type would be detected prior to release.The complaint was confirmed; however, a root cause could not be established.A capa was opened to further address this issue.A conclusion code could not be chosen as the complaint was confirmed; however, a root cause was not established.
 
Event Description
Customer alleges laryngoscope blade received broken in the package at receipt.There was no reported patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RUSCH GREEN RUSCHLITE DISP MTL MAC 4
Type of Device
LARYNGOSCOPE, RIGID
Manufacturer (Section D)
TRUPHATEK INTERNATIONAL LTD.
netanya
Manufacturer (Section G)
TRUPHATEK INTERNATIONAL LTD.
14 benny gaon street
p.o. 8051
netanya 42504 43
IS   4250443
Manufacturer Contact
katharine tarpley
3015 carrington mill blvd
morrisville, NC 27560
9194334854
MDR Report Key6044826
MDR Text Key57977996
Report Number8030121-2016-00014
Device Sequence Number1
Product Code CCW
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 10/11/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Catalogue Number004551004
Device Lot Number1604342
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/25/2016
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 10/11/2016
Initial Date FDA Received10/20/2016
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received12/01/2016
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-