MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC.(CRM-SYLMAR) MERLIN@HOME, RF TELEMETRY BASIC; PACEMAKER DATA TRANSMITTER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC.(CRM-SYLMAR) MERLIN@HOME, RF TELEMETRY BASIC; PACEMAKER DATA TRANSMITTER Back to Search Results
Model Number EX1150
Device Problems Charred (1086); Failure to Power Up (1476)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 08/10/2016
Event Type  malfunction  
Manufacturer Narrative
(b)(4).
 
Event Description
It was reported that due to a lightening strike the programmer would not boot up and the power strips were charred.
 
Manufacturer Narrative
Final analysis confirmed the reported complaint; visual inspection noted that the prongs were charred and the circuit board exhibited burn marks.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MERLIN@HOME, RF TELEMETRY BASIC
Type of Device
PACEMAKER DATA TRANSMITTER
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
jaime chavez
15900 valley view court
sylmar, CA 91342
8184934022
MDR Report Key6045222
MDR Text Key58026425
Report Number2017865-2016-06684
Device Sequence Number1
Product Code DRG
Combination Product (y/n)N
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Type of Report Initial,Followup
Report Date 08/10/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberEX1150
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/26/2016
Initial Date Manufacturer Received 08/10/2016
Initial Date FDA Received10/20/2016
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received03/29/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/06/2012
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-