Model Number N/A |
Device Problem
Fracture (1260)
|
Patient Problem
Foreign Body In Patient (2687)
|
Event Date 09/28/2016 |
Event Type
Injury
|
Manufacturer Narrative
|
Current information is insufficient to permit a valid conclusion about the cause of this event.A follow up report will be sent upon completion of the device evaluation.
|
|
Event Description
|
It was reported that the tip of a pedicle probe broke off within the patient during a t10-pelvis fusion procedure.The tip was not able to be removed from the patient; the patient retained a foreign body.
|
|
Manufacturer Narrative
|
The returned device was examined.The curved section of the tip was found to have fractured off and was not returned; the complaint is confirmed.A review of the manufacturing records did not identify any issues which would have contributed to this event.The labeling was reviewed and found to contain sufficient instructions regarding proper device usage.The cause of the event is unknown.Further event details were requested from the reporter but no further information was provided.
|
|